QUESTIONS & ANSWERS
If you have any questions regarding method validation, instrument performance and calibration, method development and pharmaceutical analysis, please forward them to cvg@cvg.ca. We will try to answer your questions or direct your questions to the experts who may be able to answer them.
I am looking for information regarding the theoretical values of caffeine UV maxima and minimum. I am trying to incorporate those values in my HPLC VW Detector SOP and I like to have something official about those values. I have looked into Merck Index, Handbook of Chemistry and Physics but none of those mention those values. Several vendors have provided information but I got values that differ somehow from each other. For example, one vendor uses 272 nm and another 273 nm, For the minimum, one vendor uses 244 nm another uses 245 nm.
The lamda maximum and minimum of a caffeine solution depend on the solvent used and the spectral bandwidth the spectrophotometer or detector. Typically, the spectral bandwidth of UV detectors is quite broad. For a caffeine solution in water measured using a spectral bandwidth of 6 nm, the lamda maximum occurs at 273 and 205 nm and minimum occurs at 244 nm. These values are taken using a calibrated high resolution spectrophotometer. I hope this help.
I would like to validate two incubators (25 oC/60% RH and 40 oC/75% RH). We are a pharmaceutical plant, so we have to meet FDA regulations. I would like to have these validations performed by a specialized firm, since we are not equipped to perform them. Can you help me with this matter?
The following companies do this type of work:

Can you please let me know, what we mean by Quantity (Q)?

Where or how is it derived and what are the advantages and disadvantages of using Quantity (Q) in dissolution testing ?
I am not sure about the origin of "Q-value" in dissolution test. However, USP defines the criteria for stage 1, 2 and 3 dissolution testing using Q-value, therefore all US submissions (e.g., NDA, ANDA) to FDA should have the dissolution specifications based on Q-values. Stage 1,2 and 3 testings are only used by USP and therefore they are the only one who use Q-values. For submissions to other countries (e.g., EU), there is no need to use Q-values.

For example, the dissolution specifications for the same drug product will be:
US submission: meet USP requirements, where Q=75% in 30 minutes. Other submissions: not less than 80% dissolved in 30 minutes. They are the same specification limit, as stage 1 USP testing requires all units to be not less than Q+5% (i.e., 80%).

Keep in mind that Q=75% means all units will be not less than 80% (not 75%) dissolved.
We are about to transfer some validated analytical methods to our own laboratory. Am I correct in thinking ruggedness and precision are the only tests required to certify us to use these methods. By ruggedness I mean 2 analysts test 6 samples from each of 2 different manufacturing lots on 2 different instruments on 2 different days?

Would the %RSD of the grand mean have to be less than 2.0% (main active assay) Would this spec apply to within and between laboratories. Is two different manufacturing lots enough? Is this approach overkill or not enough?
I am assuming the method you are transferring is the potency method. Based on the info you mentioned, your proposal would be sufficient. The specification should be applicable to both labs. The acceptance criteria for the method transfer will be based on the variability of your method and should be within the normal acceptable variability.
However, for impurity methods, you will need to ensure sensitivity of the instrument in the new lab vs. your lab.
When doing the precision test on the assay of main actives, I made six sample preparations from the same sample, I then calculated the % RSD of six replicate results. The criteria is % RSD less than 2.0%. How to perform this test on the tablet dissolution, what is the criteria in this case?

For the ruggedness test (deviation from dissolution method conditions), how many factors that I have to take into consideration, and what are the accepted % RSDs (compared to the unmodified conditions)?

If the dissolution assay is to be carried out by a second analyst (on the same HPLC and dissolution unit) what is the accepted % RSD? These questions are raised because the individual tablet weights are different, and the content uniformity test has a high accepted % RSD (for example in USP, % RSD is not more than 6.0%)
I will perform 6 replicate of tablet dissolution and analyze the results (by HPLC or whatever you choose). RSD should be less than 5%. This precision includes the variation caused by dissolution and by HPLC analysis.

I prefer to call it "robustness" instead of "ruggedness". Technically speaking, "ruggedness" is "day to day variation, that means ICH intermediate precision". The number of factors are up to you. In most cases, I will use less than 6 factors. Some of the factors that I will investigate are " degas method, rpm, sinker type, dissolution medium composition). In general, I expect to see the "mean" dissolution results from all conditions to be within 5% of each other.

In order to avoid tablet-to-tablet variation to affect your results, you can compare the "mean" of one analyst versus the "mean" of another analyst. The mean results should be less than 5% of each other. In addition, you can use some "normalization" technique to account for tablet to tablet variation. I can't go into too much details here.
I was wondering if you could help me to find information regarding the requirements on robustness of an HPLC method. I looked in the ICH guidelines and I found the document: "Validation of analytical procedures: methodology". They mention that robustness has to be evaluated but there is no limit or requirement to meet.
The following companies do this type of work:
Validation Technologies & Pharmaceutical Technologies
I would like to validate two incubators (25 oC/60% RH and 40 oC/75% RH). We are a pharmaceutical plant, so we have to meet FDA regulations. I would like to have these validations performed by a specialized firm, since we are not equipped to perform them. Can you help me with this matter?
The following companies do this type of work:
&
Can you please let me know, what we mean by Quantity (Q)? Where or how is it derived and what are the advantages and disadvantages of using Quantity (Q) in dissolution testing ?
I am not sure about the origin of "Q-value" in dissolution test. However, USP defines the criteria for stage 1, 2 and 3 dissolution testing using Q-value, therefore all US submissions (e.g., NDA, ANDA) to FDA should have the dissolution specifications based on Q-values. Stage 1,2 and 3 testings are only used by USP and therefore they are the only one who use Q-values. For submissions to other countries (e.g., EU), there is no need to use Q-values.

For example, the dissolution specifications for the same drug product will be:
US submission: meet USP requirements, where Q=75% in 30 minutes. Other submissions: not less than 80% dissolved in 30 minutes. They are the same specification limit, as stage 1 USP testing requires all units to be not less than Q+5% (i.e., 80%).

Keep in mind that Q=75% means all units will be not less than 80% (not 75%) dissolved.
We are about to transfer some validated analytical methods to our own laboratory. Am I correct in thinking ruggedness and precision are the only tests required to certify us to use these methods. By ruggedness I mean 2 analysts test 6 samples from each of 2 different manufacturing lots on 2 different instruments on 2 different days?

Would the %RSD of the grand mean have to be less than 2.0% (main active assay) Would this spec apply to within and between laboratories. Is two different manufacturing lots enough? Is this approach overkill or not enough?
I am assuming the method you are transferring is the potency method. Based on the info you mentioned, your proposal would be sufficient. The specification should be applicable to both labs. The acceptance criteria for the method transfer will be based on the variability of your method and should be within the normal acceptable variability.
However, for impurity methods, you will need to ensure sensitivity of the instrument in the new lab vs. your lab.
When doing the precision test on the assay of main actives, I made six sample preparations from the same sample, I then calculated the % RSD of six replicate results. The criteria is % RSD less than 2.0%. How to perform this test on the tablet dissolution, what is the criteria in this case?

For the ruggedness test (deviation from dissolution method conditions), how many factors that I have to take into consideration, and what are the accepted % RSDs (compared to the unmodified conditions)?

If the dissolution assay is to be carried out by a second analyst (on the same HPLC and dissolution unit) what is the accepted % RSD? These questions are raised because the individual tablet weights are different, and the content uniformity test has a high accepted % RSD (for example in USP, % RSD is not more than 6.0%)
I am assuming the method you are transferring is the potency method. Based on the info you mentioned, your proposal would be sufficient. The specification should be applicable to both labs. The acceptance criteria for the method transfer will be based on the variability of your method and should be within the normal acceptable variability.
However, for impurity methods, you will need to ensure sensitivity of the instrument in the new lab vs. your lab.
I was wondering if you could help me to find information regarding the requirements on robustness of an HPLC method. I looked in the ICH guidelines and I found the document: "Validation of analytical procedures: methodology". They mention that robustness has to be evaluated but there is no limit or requirement to meet.
The current ICH guideline on method validation (Q2B) only gives a brief outline on the expectations for method robustness. In general, the following suggestions may be considered.

Focus on critical parameters only: Since it is impossible to investigate all parameters in the method, I only focus on those critical parameters (i.e., parameters that have significant impact on the method performance). We can usually identify the critical parameters during method development. For example, if you know that % organic will cause significant change in peak shape, this may be one of the parameters in the method validation.

Variation range in method validation:
The range should reflect the expected day-to-day variation, so the range should be quite tight. For example, if % organic is one of the parameters, I will suggest to use +/- 2% as the range as this reflects the expected range when the procedure is performed properly. I have seen method robustness problems just because of using a too-wide variation range in method validation.

Investigate sample preparation/ HPLC separation independently

Method robustness should be investigated separately for sample preparation vs. HPLC chromatography. Sample preparation: look for effect of varying sample preparation on the results (e.g., sample solvent composition, sonication time....) HPLC conditions: look for similar effects by varying HPLC conditions (e.g., detection wavelength, mobile phase composition, column temperature...)

Try to use experimental design to minimize work

Since the workload will increase exponentially when you try to change multiple parameters at the same time, I like use traditional experimental design to minimize the workload (e.g., fractional factorial design...)

Tie to method procedure

If you demonstrate that the method is very sensitive to a particular parameter, then you need to specify a very tight range in method procedure (e.g., % organic should be between 48% to 52%). This range will ensure the method performance is reproducible on a day-to-day basis.

How often should the calibration weights be recalibrated? Our balances are used in laboratory and processing situations.
We calibrate the standard weights for every year, in order to have control over their accuracy as indicated by GMP's. And they also can be calibrated under request if there is any suspicion of change of weight or for any sign of deterioration. This will be my recommendation since auditors were very comfortable with this policy.
Please share some experiences on any lab glassware or dishwasher cleaning validation. The topic seems to be one of the frequently asked questions by the regulatory agencies.
For glassware cleaning validation, it may be useful to assume the worst case scenarios.

1. Select the worst case compound in the laboratory. That is the compound with the lowest solubility in most solvent types, and in most aqueous conditions.

2. Select the worst case concentrations. There are two 2 different situations:
a. When someone prepares a high concentration of this compound (e.g., for impurity assay), clean the glassware in a normal manner. Then use the same glassware to prepare a "blank" and detect if any active present in the blank.
b. Use low volume pipettes (e.g., 2 mL, 5 mL) to pipette the high concentration solutions. Clean the pipettes and similarly determine the carry over in blank. Pipettes tend to be the most difficult to clean.

3. Acceptance criteria
The worst case is that from 2a above, if there is any carry over (from insufficient cleaning of glassware) and the same glassware is used to prepare a low concentration of the same compound (e.g., potency). Relatively speaking it will cause the highest bias in the potency assay.

In 2a. The carry over in the blank should be less than 1% of potency if the same glassware is used for potency determination.

In 2b. The carry over in the blank should be less than the uncertainty in the pipette volume for a class A pipette.

Reminder: It is always necessary to "develop" the normal cleaning process.
Different detergent, wash cycle, pre-rinse, soaking of pipettes..... will affect the cleaning effectiveness significantly. Therefore before we validate the process, we have to fine tune the procedure and ensure we'll meet the criteria.

I enjoy reading your insightful articles in "Laboratory Focus" and also appreciate your initiative to bring together calibration and validation specialists so we can learn and share information. I have two questions re: validation requirements for analytical and general methods that I am hoping you might be able to clarify.

1) Do general methods (eg, pH measurement of a product solution, osmolality determination, etc) listed in USP require validation for regulatory compliance when applied specifically to product testing?

2) I have several test methods for various finished products which require validation (eg., "Formaldehyde Determination by the Regular Nash Method" for Product A, Product B, etc.). Many of these product-specific test methods are based on a (non-compendia) general test method (eg., "Formaldehyde Determination by the Regular Nash Method"). Question: Is a full validation required/necessary for each product for compliance? Or can one full validation be performed for one product and a partial validation (eg., just specificity and, perhaps, robustness) for the rest? The assumption in the latter approach is that the fully validated method can be referenced for the missing but potentially transferable validation elements since the testing conditions were practically identical.

1) Yes, general USP methods do require some form of method validation. It is because USP usually does not include enough details, and therefore chances are scientists need to supplement with some product specific details (or some local practices) to make the procedures work. One typical example is the Karl Fischer titration method. Even though it is primarily a general procedure, I have seen much time that product specific details have to be added to the procedure. For example, appropriate sample size, number of replicate, sample handling procedure, proper titration medium, data acceptance criteria, system suitability criteria. Having said that, I don't mean to have very comprehensive method validation for some simple procedures (e.g., pH). For something like pH measurement, we just need to demonstrate the accuracy (i.e., absence of matrix interference) and precision.

2) It is perfectly fine to use method validation data from other methods. Provided that these data are not affected by sample matrix (i.e., will not change from product to product).
For example, in some cases we demonstrate the linearity of the method by plotting response vs. concentration of active with no excipient. Therefore, the same linearity results can be shared among several products of the same active.
I am inquiring about what happens to the uncertainty of measurement, when one measurement device is used to calibrate another measurement device.
For example, if a reference thermometer known to have an error of measurement of say +/- one degree C, is used to calibrate a second thermometer. In this example, the second thermometer would now have an uncertainty of +/-4deg C (1deg C times 4 = 4 deg. C).
Typically, there are two factors to be considered in doing calibration, one factor is the uncertainty ratio and the other factor is the traceability. For the Uncertainty Ratio (UR), it is recommended to be 1:4, meaning that the standard is four times better than the instrument to be calibrated. For example, if the process limits are +/- 4 deg C, an instrument with an uncertainty of +/-1 deg C should be used and calibrate the instrument with a traceable standard with an uncertainty of +/- 0.25 Deg C. However, sometimes there are limitations (technological or budget) where the UR ratio of 1:2 may be justifiable on a case by case basis. An UR equals or larger than 1 is not acceptable. In addition, any chosen standard must be traceable to a recognized agency (eg. NIST, ASTM) and re-certified periodically.
I would like to begin by saying that I enjoyed the IQ/OQ/PV course very much and I hope to attend more courses involving this subject area in the near future. My question at this time is if you could recommend to me any articles that deal with or discuss the validation of computer software.
There are three books on computer system and software validation that may be useful to you.

Validation And Qualification In Analytical Laboratories, Ludwig Huber, Interpharm Press.
Validation Automated Manufacturing And Laboratory Applications
Guy Wingate, Interpharm Press.
The Survive And Thrive Guide to Computer Validation, Teri Stokes, Interpharm Press.


Copyright © CVG 2005. All Rights Reserved
Website designed by TereDesign